Follitrix™ | Analytical Growth Factor Results
Independent analytical comparison • ELISA

More measured growth factors. By the numbers.

Follitrix was evaluated against a reference GFC kit using quantitative ELISA analysis. Across all four growth-factor markers (PDGF-BB, IGF-1, EGF and VEGF-A), Follitrix showed a higher measured concentration in the tested samples.

The comparison uses the final concentration tables, excludes out-of-range values from averages, and uses the most reliable dilution for each analyte.

+41.8%
Higher measured PDGF-BB concentrationFollitrix average vs. reference GFC kit
PDGF-BBIGF-1EGFVEGF-AELISA verified4PL curve fitPDGF-BBIGF-1EGFVEGF-AELISA verified4PL curve fit
The results

Follitrix measured higher across all four markers.

Reference-kit average is shown as 100% (the white marker). The Follitrix bar shows the measured concentration relative to that reference.

PDGF-BB
+41.8%
vs. reference average
0Ref 100%141.8%
IGF-1
+36.3%
vs. reference average
0Ref 100%136.3%
EGF
+12.8%
vs. reference average
0Ref 100%112.8%
VEGF-A
+9.1%
vs. reference average
0Ref 100%109.1%

Relative concentration index

AnalyteReference = 100%FollitrixMeasured difference
PDGF-BB100.0%141.8%+41.8%
IGF-1100.0%136.3%+36.3%
EGF100.0%112.8%+12.8%
VEGF-A100.0%109.1%+9.1%
* Differences are analytical concentration deltas versus the reference-kit average; they are not claims of clinical efficacy or treatment outcomes.
Product comparison

Follitrix beside the reference.

Blurred reference GFC kit
Reference GFC kit
Manufacturer intentionally not identified
Follitrix box
Follitrix™
Autologous Growth Factor Concentrate (GFC) Kit
Inside the testing

From sample preparation to quantitative readout.

The supplied photographs document the comparative laboratory workflow, ELISA plates, sample handling and optical-density readings.

01
Sample preparation

GFC samples prepared for comparative analytical testing.

02
ELISA assay preparation

Standards, controls and samples prepared for the selected growth-factor assays.

03
Plate reading

Optical-density measurements captured from the assay plates.

04
4PL back-calculation

Four-parameter logistic standard curves used to convert corrected OD into concentration values.

Assay quality

Strong standard-curve fit.

All four tested analytes produced R² values above 0.995 for the standard curves.

0.9955EGF standard-curve R²
0.9979VEGF-A standard-curve R²
0.9997PDGF-BB standard-curve R²
4 / 4analytes with R² > 0.995
Live illustration of an ELISA 96-well colour development.
Laboratory evidence

Testing documentation.

Selected photographs from the testing process. Click any image to view it larger.